1 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
1803M Examination Couch
User's Manual
2 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
Index
1. Introduction 3
2. About this Product 3
3. Definitions 4
4. Safety 4
5. Environmental Specifications 5
6. Installation 6
7. Add-On Accessories 8
8. Operation 8
9. Maintenance & Cleaning 9
10. Warranty & Service 11
3 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
INTRODUCTION
About this User Manual
This user manual contains instructions for the proper and safe use of the examination couch, and
for patient & user safety and comfort, which has to be strictly followed.
This User Manual has to be read and followed by every person using or handling this examination
couch. In case of any query, Kindly contact Midmark India Pvt. Ltd.
ABOUT THIS PRODUCT
This Health Check Single Top Examination Couch wit 'U' cut is suitable for use in Gynaec OPD
and Exam Room.
The BP Apparatus Tray is adjustable for convenient positioning of BP Apparatus. This comes
with drawers and doors for storage and has an inbuilt sliding step stool for easy
mounting/dismounting of patients when required.
U-Cut is provided for lithotomic to help the use of speculum and forceps. Lithotomy pole
covered with soft rubber is provided for the lithotomic position.
Examination Couch Features
Please refer to the 1803M-1808 catalogue for the salient features of these products.
Technical Data
Please refer to the 1803M-1808 catalogue for the technical data of this product.
Add On Accessories
Please refer to the 1803M-1808 catalogue for various accessories available and compatible
with this product.
4 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
DEFINITIONS
Operator
The operator is the person appointed by the owner (i.e. clinic, hospital management. etc.) who will
be responsible for the safe operation of this examination couch.
User
A user is a person who is capable of assessing the clinical condition of the patient e.g. specialist
medical personnel, doctors, nursing staff, attendants, etc. Based on their training experience or
through the instructions, they are entitled to operate the examination couch on their responsibility
or to carry out work on it and can recognize and control possible risks to the patient as well as the
examination couch.
Patient
A patient is a person who is ill, infirm, disabled, or in need of care & occupying this examination
couch. Each time the examination couch is allotted to a new patient it is recommended that the
patient is instructed by the operator or the user about the functions important for his/her comfort
and safety.
SAFETY
In this User Manual, the following safety symbols are used
Danger: This symbol will appear wherever safety instructions are designed to protect people from
physical harm. The symbol stands for imminent danger of death or serious injury.
Caution: This symbol will appear where situations are described, which might be dangerous, and
which might inflict slight injuries.
Advice: This symbol will appear in front of additional helpful pieces of advice.
A dot in front of the text implies the action that is required to be taken.
5 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
Safety Instructions
Always use original accessories that have been designed and approved for use with the
examination couch
Ensure that the examination couch is assembled and installed following the instructions
given in this user manual
Restrict child visitors without proper attendants. Supervision is necessary to avoid any
mishap due to unintended movement of the examination couch
Keep some gap between walls and bed while in a stationary position and also while moving
from one place to another
Keep the user manual handy with the operator/user for ready reference
The patient should be advised not to sit on edge of any section when it is in extended
full-length position
ENVIRONMENTAL SPECIFICATIONS
Usage, Storage and Transport
Temperature -1O° C to 50°C
Relative Humidity 0% to 95% non-condensing
Atmospheric Pressure 800 to 1060 hPa
Altitude < 200Qm
6 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
INSTALLATION
1. Unpack all the components carefully.
2. lf the packing is drenched by rain, water, or other liquids, inform this matter to the
operator.
3. Each Examination couch is supplied with the following components as per the packing list:
Couch with fixed upholstery top with drawers & lockers - 1No-
Synthetic Rubber padded S.S. Lithotomy Rods with Rexine Ankle Straps - 2 Nos.
BP Apparatus Holder with fixing bracket & star knob - 1No.
Hardware -1 Packet
Notify if any damage has occurred during transportation either to the authorized agent
or directly to Midmark India Pvt. Ltd.
Installation of the examination couch should be done only by Midmark India trained
technicians
Disposal of the packing material waste following the local waste disposal regulations
prevailing
7 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
1. Insert the Lithotomy poles into its slots & tighten the same with the SS bolts
2. Insert the BP Tray Holder Rod into the bracket & tighten the same with the help of
star knob
8 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
ADD-ON ACCESSORIES
1. Paper Roll Holder (0175)
Paper roll holder to be fitted at head side & leg side
LOAD BEARING CAPACITY
Do not exceed the safe working load mentioned above
Operation
The Lithotomy Rods can be adjusted as per the desired height & can be locked with the help of
SS locks provided
The BP apparatus tray can be adjusted to the desired position with the help of star knob
The sliding step stool can be pulled out for use & can be pushed back in
This examination couch should be operated only by an authorized and trained person who
knows its operation
The patient/user should not mount or sit or place any heavy load on the backrest
section/leg section as it may cause toppling.
With due consideration to environmental limitation, materials used to make the device
are not affected by contact with any biochemicals or biowaste when used as per its
intended purpose, all materials used in the product and accessories do not facilitate
breeding or growth of bio-contaminants on their own.
= 135 kgs
= 170 kgs
9 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
MAINTENANCE & CLEANING
Maintenance
Each time before putting the Examination couch into service the user must check that
All the functions are in satisfactory working condition before the examination couch are
allotted to the next patient
Lubrication of all the mechanical joints/pins / moving parts / tightening of the bolts should
be carried out at least once in three months, to avoid unnecessary sound and wear & tear of
the examination couch
At regular intervals carry out a visual inspection and check for any mechanical damage (e.g
loose I missing rivets, screws, etc.)
If any damage or malfunction is observed, immediately report this matter to the operator
If the examination couch is not in use, it should be stored in a dry and rodent free indoor
environment on a flat surface with a gradient not more than 5 degrees
If the examination couch is not in use for a long term, it is necessary to conduct a test to
check for functioning and any mechanical damage that may have been caused
Depending on the usage, certain moving parts may require replacement due to wear & tear.
We recommend timely replacement with original spares for safety and trouble-free
operation
Mattress impermeability could be affected by needle sticks. Users should be instructed to
avoid punctures caused by improper use of needle sticks. Failure to do so could result in
cross-infection & patient injury. The upholstery should be regularly inspected for punctures,
rips, tears, or other such damage & replaced with the suitable original mattress if required.
Smoking should be prohibited near the upholstery with/without patient, as this is a
fire hazard, and mattress, sheets & pillows generally do not have flame retardant
properties.
10 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
Cleaning & Disinfecting
The Examination couch is not suitable for machine wash or steam cleaning
Do not use water jet for cleaning
The Examination couch cannot tolerate the influence of strong solvent, basic or alkaline
liquids & caustic agents
The cleaning agents must not contain any substances which bring about any change in the
structure or adhesive property of the plastic
The examination couch should be properly cleaned and disinfected to avoid infection and
stains before allotment to a new patient.
We recommend the use of approved disinfectant and cleaning agents to clean the
examination couch. Take extra care to clean areas that may trap dust or dirt. All detachable
accessories and its component can be cleaned separately.
Do not expose the Examination Couch & Table to excessive moisture as corrosion may cause
personal injury or equipment damage.
Disposal:
Product and its defective parts especially elements made of plastic, steel, rubber material
should be disposed of following the statutory rules and regulations prevailing for such
disposal. The user is advised to contact Midmarkindia Pvt. Ltd. for further assistance.
11 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
Routine Inspection Check List: Inspection by the User (Recommended Frequency: Every new
allotment but not later than three months)
Checks
Observation
Ok
Not ok
Visual check of the mechanical components
Main Body
Upholstery
Damage, Deformation
Upholstery
Knobs, Screws and
rivets
Performance check
of all mechanical
components
Detachable lithotomy
Pole
BP Apparatus Tray
Sliding Step Stool
Smooth
Locking/unlocking
Damage, Torn, Soiled
Loosened, Missing
Secure retention
Secure Retention
Smooth Movement
Drawer / Door Lock
Sign
Result of inspection
Date
12 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
Troubleshooting Guide
Problem
Possible Causes
Remedy
Couch / Table shaking
or uneven
• Check the Floor level &
legs
• Adjust / replace Nylon
Bush
Headrest not working
• Check Alignment of Gas
Spring
• Gas Spring defective /
damaged
• Readjust the Gas Spring
Bracket &
Cover
• Replace the Gas Spring
13 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
WARRANTY & SERVICE
Midmark India Pvt. Ltd. warrants solely to the end-users of this Examination couch that for a period
specified in the warranty certificate, such Examination couch shall perform as per its intended use.
The foregoing warranties shall not apply if the Product has been subjected to:(a) physical abuse,
misuse, abnormal use, or use not consistent with Midmark India Pvt. Ltd.'s published directions, (b)
any modification by anyone other than Midmark India Pvt. Ltd., or third parties designated by
Midmark India Pvt. Ltd., or (c) fraud, tampering, negligence or accidents, depending upon the nature
of the products.
Limitation Of Liability: In no event shall Midmark India Pvt. Ltd., be liable or obligated to the end-user
in any manner for any special, non-compensatory, consequential, indirect, incidental, statutory or
punitive damages of any kind. Including, without limitation. For injury or loss of life, lost profits lost
sales. Lost revenue, loss of any software or data. or loss of use of any hardware, software or data,
regardless of the form of action, whether in contract, tort, negligence, strict product liability, or
otherwise, even if Midmark India Pvt. Ltd., has been informed of or is aware of the possibility of any
such damages in advance. Midmark India Pvt. Ltd.’s total aggregate liability under this agreement
and in relation to anything which Midmark India Pvt. Ltd., has done or not done in connection with
this agreement shall be limited to the price of the products purchased and paid for by the end-user
during the period in which such liability arises. The limitations set forth above shall be deemed to
apply to the maximum extent permitted by applicable law and notwithstanding the failure of the
essential purpose of any limited remedies.
In order to maintain functional safety and right to claim under the warranty only original 'Midmark'
replacements parts may be used for ordering replacement parts, customer service request and
further question, please contact our technical support department.
14 | P a g e TF-NMF-1803M-Usermanual: Rev.No.01: 19.05.20
Disclaimer
Every effort has been made to ensure that the information in this user manual is valid. Midmark
India Pvt. Ltd., reserves the right to change the information in this document without prior notice."
MANUFACTURER’S ADDRESS
Corporate Office:
Midmark India Pvt. Ltd.
Art Guild House, ”A” Wing, Unit No.15 & 16, Lower Ground Floor, Phoenix Market City, LBS
Road, Kurla(West), Mumbai, 400070
Toll Free: 1800 22 8020 | Tel: .:+91 22 4915 3000
Registered Office:
Midmark India Pvt. Ltd.
Janak House, Opp. IOC Depot, Shaikh Misry Road, Wadala (E),
Mumbai 400037, India
Tel: + 91 22 2413 0407
Toll-Free No.: 1800 22 8020 | Email: technicalsupport@midmark.com
MIDMARK INDIA PVT LTD.
Art Guild House, A-Wing, Unit No. 15 & 16, Phoenix Market City, LBS Road, Kurla (W),
Mumbai - 400 070 Toll Free: 1800 22 8020 | Tel: +91 22 4915 3000 | Fax: +91 22 4915
3100 www.midmark.in | india@midmark.com
Improving caregiver and patient experience
Fixed height on examination table may restrict patient’s access and impact the healthcare delivery.
Why is height adjustment in examination table recommended?*
• 44% of practice administrators reported that portions of an examination were omitted when faced with
barriers during the assessment of patients with disabilities.
• 76% of practice administrators indicated that patients were examined in wheelchairs when they
could not transfer themselves onto an exam table.
The JX3000 features adjustable height for easy access, ensuring comfort for all patients, including those with
limited mobility. Its stable design supports smooth positioning, including Trendelenburg and Reverse
Trendelenburg options, enhancing examination efciency. Castors ensure easy mobility, making it practical
for any clinical setting.
JX3000
JX3000
Exam Room Solutions
Upholstery Colour Options
Castors allow for easy relocation of
JX3000
Castors
To minimize surface contamination
Paper roll holder
JX3000:06:12:08:24
Midmark (India) Private Limited
Head Ofce: Art Guild House, FOF, B-16, "B" Wing, 4
th
Floor, Phoenix Market City, LBS Road, Kurla (West), Mumbai - 400 070.
Toll Free: 1800 22 80201Tel: +91 22 4915 3000 Fax: +91 22 4915 3100 I www.midmark.in india@midmark.com
Overall dimensions: 2040 mm x 660 mm L x W
Minimum height: 505 mm, Maximum height: 855 mm
Trendelenburg & reverse trendelenburg: 19°
Upholstered platform width: 600 mm
Back rest length: 630 mm
40 mm diameter castors
Motorized height, trendlenburg and reverse trendelenburg positions
operated by handset
Backrest adjustable over gas spring mechanism
Back support angle: 0°- 65°
Safe working load: 170 kg
Two section xed rexine upholstered top 65 mm thick
Antibacterial, antifungal and re retardant upholstery rexine
Technical Specifications
Add-On Accessories:
Standard Features:
Power in (Mains Power Supply): 100 V to 240 V 50/60 Hz, Max 2. A.
Electrical Shock Protection: Class 1
Degree of Shock Protection: Type B
Liquid Ingress Protection: IPX6
Duty Cycle: 10% (Two minutes for every eighteen minutes)
Electrical Parameters
Advanced Patient Positioning Features
Safety Side Rails BP apparatus tray
Height-adjustable BP tray
Elevated backrest
Maximum height
Minimum height Trendelenburg
Reverse Trendelenburg
Moss Navy
Raisin Grape
Shadow
Pista Healing Waters Robust Brown
*The Exam Chairs Impact on Patients and Providers
//www.midmark.com/medical/resourse-library/blog/the-exam-chair’s-impact-on-patients-and-providers
Scan here to access the digital
copy of this yer
Safety Standards
General safety for
medical electrical equipment
General safety for
medical equipment
EMI/EMC
IEC 60601-1-1
IEC 60601-2-52
IEC 60601-1-2
//www.midmark.com/medical/resource-library/blog/the-exam-chair’s-impact-on-patients-and-providers
*The Exam Chair’s impact on Patients and Providers
Battery backup for
uninterrupted power supply
*± 10 mm Engineering Variation ** ± 2° Engineering Variation
When couch is congured with standard accessories
<
Single piece can either be installed on
either side of the couch